55
Direct current
Range Accuracy Resolution:
VC265 VC280
400 µA
±(1.3% + 3) ±(1.3% + 3)
0.1 µA
4000 µA 0.001 mA
40 mA
±(1.5% + 5) ±(1.5% + 5)
0.01 mA
400 mA 0.1 mA
4 A
±(1.8% + 8)
±(4.6% + 39) 0.001 A
10 A - - - - 0.01 A
40 A - - - - ±(4.6% + 7) 0.01 A
Overload protection: Fuses; Overload protection;
600 V; measuring time limit mA-PTC-Fuse 600 V;
>5 A: max. 10 s with break Measuring time: unlimited
of 15 min
Alternating current
Range Accuracy Resolution:
VC265 VC280
400 µA
±(1,6% + 3) ±(1.6% + 3)
0.1 µA
4000 µA 0.001 mA
40 mA
±(2,0% + 5) ±(2,0% + 5)
0.01 mA
400 mA 0.1 mA
4 A
±(2,0% + 8)
±(4.6% + 39) 0.001 A
10 A - - - - 0.01 A
40 A - - - - ±(4.6% + 7) 0.01 A
Frequency range 45 Hz - 1 kHz;
Frequency range
TrueRMS; Crest factor max. 3.0
40 Hz - 400Hz; Crest factor
Overload protection: Fuses600V;
max. 3.0 Overload
Measuring time limited >5 A: protection; mA- PPTC
max. 10 s with break of 15 min
Fuse 600 V